Del Mar Ventures> Product List> MCP products> MCP Applications>
In particular, the MCP detector is highly sensitive to the biologically
important but previously difficult to detect low energy (average 6~keV, endpoint
18.6~keV) beta emission from tritium. We report 3H sensitivities and
linearity derived from images of a 3H standard, together with images
of 3H-Putrescine doped semi-issue sections of rat lung and isolated
single cells from rabbit lung. We compare these results with those of previous
attemps to digitally image tritium.
The present approach [9] utilises the high electron detection efficiency of
MCPs [10,11] for the direct imaging of the betas from samples placed in close
proximity to the input MCP in vacuo (10-6mbar).
The main source of detector background in "standard" lead silicate glass MCPs is from the 40K beta-emission in the potassium, a major glass consituent [14]. Low noise MCPs, however, are manufactured from proprietary potassium and rubidium-free glass. All detector body parts were manufactured from PCTFE [15], rather than such materials as Macor, which itself contains potassium [16]. Dark noise rates for both the present detectors were 0.12 counts cm-2s-1 above a discriminator level of 0.05 times the peak detector gain [13].
The ultimate sensitivity of any beta detector depends, for a given electron detection efficiency, on reducing the background noise to a minimum. As most of the residual low noise detector background is due to 1.46 MeV 40K gamma rays from concrete in the laboratory walls and floor, the detector background can be reduced to cosmic ray levels (0.015 counts cm-2 s-1 [17]) by surrounding the detector with lead shielding [13].
| Parameter | Value |
|---|---|
| Thickness (L) | 1.5mm |
| Channel diameter (D) | 12.5 microns |
| Channel Pitch (p) | 15.0 microns |
| L:D | 120:1 |
| Channel Bias Angle | 6° |
The present configuration of our detector, designed for X-ray detection, operates with a potential of -4.5kV on the front plate, effectively repelling a large fraction of the the low energy 3H betas. We would therefore anticipate a factor of two increase in beta count rate were the detector operated with zero potential on the input MCP, and with the readout element at high positive potential.
The spatial resolution was estimated, from the edge reponse function [19] of the brightest cell, to be 400 microns FWHM for the tritium-MCP separation of 0.5mm. However, the intrinsic resolution of the large area detector, as previously determined from X-ray measurements, was ~80 microns [13]. Spatial resolution of detectors using resistive anode readout scales inversely with resistive anode side length. Improved imaging resolution for 3H betas can therefore be achieved by (a) reducing the size of the detector and (b) bringing the sample closer to the surface of the input MCP (below). The elimination of the retarding input electric field would also serve to improve resolution.
Figure 2 shows the measured linearity of the large MCP detector for 3H betas and is derived from the cells of the tritium standard. The dashed horizontal line represents the background noise level. The maximum count rate of the brightest cell corresponds to a count rate per channel of only 7.4x10-5s-1. Fraser et al [20] have shown that MCP detector linearity is preserved up to a count rate of order 0.1 channel-1s-1. Thus, the three orders of linearity demonstrated in figure 2 extends in principle to six orders of dynamic range.
The sample mounted on a glass cover slip was placed in direct contact
with the MCP input surface. Figure 3(a) shows the image from a lung tissue slice
after 76 hours. Comparison with an optical light image (figure 3(b)) obtained
using a binocular microscope, shows that we can image the gross features of the
sample. The air passages, both perpendicular and in the plane of the sample,
where there is no uptake of the radiolabel are clearly visible. Tritium is known
to concentrate in certain cell types (< 10 micron in size) [21] which, due to
the limitations of the present detector resolution, we could not resolve. We
have estimated, from figure 3(a), an upper limit to the image resolution of the
small format detector of 100 microns since the edge of the sample, from optical
microscope inspection, is only very approximately straight.
The activity of the tissue sample under test was unknown but from the known MCP electron efficiencies (50%), the measured noise subtracted sample count rate (0.059 counts s-1 sample thickness (1 micron) and an assumed sample density of 1g cm-3 we have estimated the sample activity to be 0.52 micro-Ci per gram.
A second sample (isolated macrophage cells from a rabbit lung, not shown) was
examined with the small detector, having been labelled, as before, with tritated
Putrescine (following the method of Dinsdale et al. [21]). The sample was
imaged once, then the detector housing was surrounded by lead (estimated solid
angle coverage, 35%) and a second image accumulated. Dark noise measured on the
shielded detector was a factor of ~0.3 lower than the unshielded value.
Signal-to-noise ratios were 7.64:1 and 10.88:1 for the unshielded and shielded
configurations respectively, confirming the improvement in sensitivity predicted
above.
Gas proportional counters can detect tritium betas with essentially 100% efficiency only if the sample is "built-in" into the detector [2,3]. The optical avalanche chamber described by Tribollet et al. [3] exhibits a count rate, for a given activity, a factor of 3.5 higher than our MCP detector in its present (electrically unfavourable) configuration (see Results: Tritium Standard). The MCP detector however, has a much better intrinsic spatial resolution compared to ~500 microns given for the avalanche chamber.
Images plates with specially reduced dead layers can be used for imaging tritium [7]. One manufacturer [22] cites a minimum detectable activity for 3H of 1.67 Bq/mm2/hour which is a factor of ~80 poorer than our measured minimum count rate (0.001 mm-2s-1). Spatial resolution for an image plate is ~150 microns FWHM, comparable to that of the present detector.
A recent paper [4] describes the use of avalanche photodiodes (APDs) for measuring the tritium content of biological specimens. The detection efficiency of such devices are limited by the thickness of the surface dead layer which acts as a barrier to the low energy tritium betas. Gordon et al. reports tritium beta efficiencies of 27% for devices having large 1mm2 pixels.
Our measurements of (a) 35S-labelled DNA electrophoresis gels and
(b) 14C-labelled rat whole body tissue slices will be described in a
future report.
Original text:
http://www.src.le.ac.uk/projects/bioimaging/publications/direct.html
Del Mar Ventures> Product List> MCP products> MCP Applications>